Determination of N-nitrosamines in Water by Gas Chromatography
Coupled with Electron Impact Ionization Tandem Mass Spectrometry |
Ki-chang Lee1, Jae-hyung Park1, Wontae Lee2 |
1Gyeongsangbuk-do Institute of Health and Environment 2Kumoh National Institute of Technology |
Corresponding author |
Wontae Lee ,Tel: 054-478-7636, Fax: 054-478-7859, Email: wtlee@kumoh.ac.kr
|
|
Received: October 29, 2014; Revised: November 13, 2014; Accepted: November 19, 2014. Published online: November 30, 2014. |
|
ABSTRACT |
This study assessed analysis of N-nitrosamines by separation, identification, and quantification using a gas chromatography
(GC) mass spectrometer (MS) with electron impact (EI) mode. Samples were pretreated by a automated solid phase extraction (SPE)
and a nitrogen concentration technique to detect low concentration ranges. The analysis results by EI-GC/MS (SIM) and EI-GC/
MS/MS (MRM) on standard samples with no pretreatment exhibited similar results. On the other hand, the analysis of pretreated
samples at low concentrations (i.e. ng/L levels) were not reliable with a EI-GC/MS due to the interferences from impurity peaks.
The method detection limits of eight (8) N-nitrosamines by EI-GC/MS/MS analysis ranged from 0.76 to 2.09 ng/L, and the limits
of quantification ranged from 2.41 to 6.65 ng/L. The precision and accuracy of the method were evaluated using spiked samples at
concentrations of 10, 20 and 100 ng/L. The precision were 1.2~13.6%, and the accuracy were 80.4~121.8%. The R2 of the
calibration curves were greater than 0.999. The recovery rates for various environmental samples were evaluated with a surrogate
material (NDPA-d14) and ranged 86.2~122.3%. Thus, this method can be used to determine low (ng/L) levels of N-nitrosamines
in water samples. |
Key Words:
N-Nitrosamines, Solid Phase Extraction, EI-GC/MS/MS, Method Detection Limit, Recovery |
|